Specifications, Available input ranges, Accuracy – Fluke 700G User Manual
Page 12: Media compatibility, Environmental, Mechanical specifications

700G Series
Calibration Manual
4
Specifications
Available Input Ranges
See PI Ranges and Resolution for available ranges in psi plus equivalent ranges and
resolution for all engineering units.
Accuracy
Positive Pressure ....................................................... ±0.05 % FS
Vacuum .................................................................... ±0.1 % FS
Temperature Compensation ..................................... 15 °C to 35 °C (59 °F to 95 °F) to
rated accuracy
Note: For temperatures from -10 °C to 15 °C and 35 °C to 55 °C, add .003 % FS/°C
Media Compatibility
15, 30 psi .................................................................. any clean dry non-corrosive gas
100, 300, 500, 1000 psi ............................................ any liquids or gases compatible with
316 stainless steel
Above 1000 psi ......................................................... any non-flammable, non-toxic, non-
explosive, non-oxidizing liquid or gas
compatible with 316 stainless steel.
Environmental
Operating Temperature ............................................. -10 °C to +55 °C (14 °F to 131 °F)
Storage ..................................................................... -20 °C to +70 °C (-4 °F to +158 °F)
Humidity .................................................................. 10 % to 95 % RH Non-condensing
Pollution Degree ...................................................... 2
Altitude ..................................................................... 3000 m maximum operating altitude
Agency Approvals .................................................... P, , ), (
Mechanical Specifications
Dimensions .............................................................. 11.4 x 12.7 (cm), depth = 3.7 cm
(4.5 x 5 (in), depth= 1.5 in)
Pressure
Connection ....................................................... ¼ in NPT Male
Housing ............................................................ Cast ZNAL
Display
5-1/2 Digits, 16.53 mm (0.65 in) high
20-Segment bar graph, 0 to 100 %
Power
Battery ............................................................. three size AA alkaline batteries
Battery Life ..................................................... 1,500 hours without backlight
(continuous on), 2,000 hours at slow
sample rate