ITC Hemochron Response Whole Blood Coagulation System User Manual

Page 10

background image

10

Selecting a Different Case or Modifying the PID for the Selected Case

1.

Display the Patient Info screen:

2.

Press 1 to display the screen for entering PID. Enter the desired PID, and then press YES.

3.

If the entered PID does not correspond to a stored case, the user is prompted whether a new case is
to be created:

Note: If a PID for an open case is entered, that case is selected and the Patient Info screen is
displayed. If a PID for a closed case is entered, the Closed Case menu is displayed to allow the
operator to either re-use patient information from that case or delete the case (page 19).

4.

Press YES if a new case is to be created. Otherwise, press NO to display a prompt to modify the case

PID:

5.

Enter the new PID, and then press YES.

Entering or Revising Patient Information

1.

Display the Patient Info screen:

2.

Press 2 to display the Enter Height screen:

or

3.

Use the numeral keys to enter the patient height (from 0.3 to 2.5 meters, or from 1 foot, 0 inches to
8 feet, 3 inches, depending upon the units chosen). If a decimal point is used, it is automatically

entered.

4.

Press YES to store the entered height and display it on the Patient Info screen.

Note: If an incorrect entry is made, press CANCEL, NO or a MENU key to cancel the entry
and redisplay the Patient Info screen.

5.

Press 3 to display the Enter Weight screen:

or