8.4 Maintaining the water circulation
In the case of discontinuous use with tap water
(single measurements with long gaps) a stabiliser
must be added to the water circulation to prevent
the formation of algae.
Add approx. 4 ml IKA
®
-Aqua-Pro C 5003.1 to the
tank (see Accessories).
Switch the pump on via the service menu (sec-
tion “Pump“). Switch the pump off again after
30 - 60 s.
You can also add the 4 ml Aqua-Pro to the water
maintained at a constant temperature in the last
measurement (before long gaps in measuring).
If the appliance is not going to be in opera-
tion for a long period of time, it is advisable
to completely empty the calorimeter's water
circulation.
The water must be drained out before trans-
portation.
Empty the outer vessel via the service menu (sec-
tion 7.2 “Outer vessel“). The emptying hose must
be locked in place in the plug-in coupling (sec-
tion 4.2).
Empty the tank by locking the emptying hose in
place in the plug-in coupling (section 4.2). The
tank will empty automatically.
Press the locking knob on the plug-in coupling
to remove the hose.
8.5 Decomposition vessels
Please see the operating instructions for
C 5010 for information on decomposition
vessel maintenance!
8.6 Cleaning information
Only clean IKA
®
appliances using these IKA
®
ap-
proved cleaning agents:
Dirt
Cleaning agent
Dyes Isopropanol
Building materials
Water containing
detergent,
Isopropanol
Cosmetics
Water containing
detergent,
Isopropanol
Food
Water containing
detergent,
Fuels
Water containing
detergent,
Other materials
Please consult IKA
®
• For cleaning the micro fi lter you must empty
the outer vessel (see section 7.2).
• Screw the cover of the micro fi lter down and
pull you out the fi lter element.
• Clean the fi lter element with clear water.
• Insert the fi lter element after cleaning with the
opening inward again into the housing of the
micro fi lter and screw you on by hand the co-
ver.
• Fill the outer vessel again with water (see sec-
tion 4.7).
• Control the tightness after you the equipment
again fi lled (see section 6.6 “Fill time exceeded).