6.3 Preparing the measurement
The calorimeter is in ”waiting” mode.
Selecting MEASUREMENT (F2) will take you
to the "Prepare measurement" menu.
Enter the noted weighted sample in the
marked "Einwaage" (weighted sample) field with
an accuracy of 0.0001 g using the keyboard.
You can access all the other options by pressing
UP/DOWN (F2):
CALIBRATION
Enter "1" here to perform a calibration.
Check the other presettings:
DECOMPOSITION VESSEL
Enter the number of the decomposition vessel
used for this test. If you are only using one de-
composition vessel, you can leave the "1" in the
"decomposition vessel" fi eld.
QEXTERNAL1
Correction value for the heat energy generated
by the cotton thread ignition aid. A default value
of 50 J is specifi ed here. The default value for the
combustible crucible option is 0 J. If you use a
burning aid other than the C 710.4 IKA
®
-cotton
thread, change this value.
Place the decomposition vessel into the inner
vessel of calorimeter C 200. The decomposition
vessel must be placed between the three locating
bolts.
Pour approx. 2 l tap water maintained at a con-
stant temperature into the tank using the measu-
ring cup. Keep an eye on the level indicator!
Attaching
the ignition
adaptor
Detaching
the ignition
adaptor
1.
2.
Locating bolt
Magnetic stirring bar