Heading1 - 4.1 power options, Chapter 4, Elec – In-Situ Con TROLL PRO Installation Manual (purchased before 10 May 2012) User Manual
Page 17: 1 power options, 2 electro-static discharge (esd) recommendations, 3 customer-supplied electrical equipment, Chapter 4 electrical connections
Electrical Connections
17
Chapter 4
Electrical Connections
Danger
Only properly trained and qualified personnel should
install the Con TROLL PRO instrument described in this
manual. This instrument should be installed for use in non-
hazardous locations only.
Danger
L'installation de l'appareil “Con TROLL” décrite dans le
guide d'utilisateur doit impérativement être réalisée par
des personnes qualifiées. L'installation de cet appareil est
seulement prescrite pour des emplacements sans risques.
4.1
POWER OPTIONS
The Con TROLL PRO process controller module is available with two power
options:
• DC-powered, which requires a 9-36 VDC supply as the primary power
source. This model can trigger low voltage (<50 V) relays only (see “Input/
Output Connections” on page 23).
• AC-powered, which requires a 100-240 VAC as the primary power
source. This model can trigger low voltage relays (<50 V) through the I/O
board only (see “Input/Output Connections” on page 23) and high voltage
relays (>50 V) through the AC board only.
4.2
ELECTRO-STATIC DISCHARGE (ESD) RECOMMENDATIONS
• Before making wiring connections, discharge any static electricity from
your body before touching circuit boards and other internal components
by touching a grounded metal object to remove any charge from your
body.
• When making wiring connections, make sure to remain properly grounded
by wearing an ESD wrist strap or similar device.
4.3
CUSTOMER-SUPPLIED ELECTRICAL EQUIPMENT
The end user must supply the following:
• 18-12 AWG wire for electrical connections using conduit or 100-240 VAC
power supply
• 24-12 AWG cable for connecting relays and PLC devices
• 2 alkaline C-cell batteries (optional—for backup purposes only)