beautypg.com

Brooks, Model 5851e, Caution – Brooks Instrument 5851E User Manual

Page 38

background image

4-6

Brooks

®

Model 5851E

Section 4 Maintenance

Installation and Operation Manual

X-TMF-5851E-MFC-eng

Part Number: 541B104AAG

November, 2008

When performing the following procedure the valve stem must be
removed without cocking it to prevent damage to the valve spring.

CAUTION

4. Carefully remove the valve stem assembly (6).
5. Remove the plunger assembly (7, 8, 9, 11).
6. Remove and note the position of the valve spring spacers (10), which

may be located above and/or below the lower valve spring (8).

7. Unscrew the orifice (12) from the flow controller body (14) using the

orifice removal tool (See Section 5, Table 5-2).

8. Remove the three screws (20) attaching the electronics cover. Remove

the electronics cover (23).

Be careful not to stress the sensor lead wire to sensor assembly
junction when removing the sensor connector from the PC Board.
If the sensor lead wires are stressed an open in the sensor wiring
could result.

CAUTION

9. Unplug the sensor connector from the PC Board. Remove the two

screws securing the bracket (24) and PC Board (15). Remove the
bracket and PC Board.

10.Remove the two allen nuts using an 1/8" allen wrench (18) and washers

(19) securing the sensor assembly (16). Remove the sensor assembly.

Note: Do not attempt to disassemble the sensor assembly.

Do not scratch the O-ring sealing surface.

CAUTION

11.Remove the sensor assembly O-rings (17) from the flow controller body

(14). Using the Brooks O-ring removal tool will help prevent scratching
the sealing surface.

12.Remove the adapter fittings (27) from the flow controller body (14).
13.Remove the 4 screws (31) from the end block (30) and carefully

remove the end block.

14.Remove the restrictor assembly (21) from the inlet side of the flow

controller body (14) using the header removal tool (Section 5, Table 5-
2). Note the rotational position of the restrictor assembly. (Refer to
Section 4-7).

B. ASSEMBLY

Do not get Halocarbon lubricant on the restrictor element (21) or
hands. Rubber gloves are recommended. This is a special inert
lubricant which is not easily removed.

CAUTION