7 MAINTENANCE AND TROUBLESHOOTING

Invacare®MA60 Series

32

Part No 1148137

8.

Perform the following steps to clean the mattress cover:
A. Using a clean sponge or paper towel, wipe down the cover surface with a dilute detergent solution of quaternary cleaner disinfectant or

other germicidal detergent solution.

B.

Remove the cover and launder it using the following method:

C. Place the cover in a washing machine.
D. Wash with warm water (below 120°F).
E.

Add detergent and disinfectant according to the manufacturer’s instructions.

F.

Remove excess water.

G. Set the dryer to the lowest setting (below 120°F).
H. Dry the cover until it is completely dry.

9.

Perform the following steps to clean the control unit and hose fittings:
A. Wipe all controls, chassis and hose fittings with a quaternary disinfectant solution.
B.

Using a nylon brush, gently clean all crevices as they can harbor microorganisms.

C. Air dry all treated surfaces.

10. Perform the following steps to clean the mattress components (air cushions, mattress base, etc.):

A. Using a clean sponge or paper towel, wipe down the mattress components with a dilute detergent solution of quaternary cleaner disinfectant

or other germicidal detergent solution.

B.

Wipe the mattress components with a clean dry cloth.

ƽ

CAUTION

Laundry workers should always wash their hands before working with clean bedding.
DO NOT overload the machine.
DO NOT use chlorine bleach because it may damage the fabric coating.
High air temperatures will damage the fabrics and void the Invacare warranty.