5 GENERAL GUIDELINES
11
Installation
ƽ
WARNING
The MA60 and MA65 systems (including bariatric and raised side rail) is recommended to be installed on medical bed
frames with bed side or assist rails. It is preferred that the rails be in the raised position whenever a patient is on the bed.
Health care professionals assigned to each case should make the final determination whether side or assist rails are
warranted after assessing patient risks of entrapment and falls in accordance with State patient restraint legislation or facility
interpretation of such legislation.
Controls on the footboard may be obstructed by the control unit on a few bed frames. It may be necessary to relocate the
control unit. Refer to Installing the Control Unit on page 18.
Check that air hoses and power cord are clear of the moving bed components before placing a patient on the bed. Operate
all motorized functions through their full range of motion to be certain that there is no pulling, interference or pinching.