6 operation – Invacare MicroAir MA60 User Manual
Page 25

6 OPERATION
25
5.
Perform one of the following:
•
Cancel - Press the Fowler button to exit Fowler Transmitter Setup without linking to a transmitter.
•
Save - Wait 60 seconds without pressing a key.
Max Inflate/Low Air Loss Button
MA60 Series
1.
Press the Max Inflate button (
) to select the max inflate mode.
•
Max Inflate Mode - In this mode, the mattress inflates rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound
every three minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the Max Inflate mode deactivates and the control
unit defaults to the previous setting.
MA65 Series
1.
Press the Max Inflate/Low Air Loss button (
) to select the max inflate mode or the low air loss mode.
•
Max Inflate Mode - In this mode, the mattress inflates rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound
every three minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the Max Inflate mode deactivates and the control
unit defaults to the previous setting.
It takes 5-10 minutes for the mattress to inflate fully (inflation time depends on size of mattress).
Max Inflate mode can be manually disengaged by pressing the Max Inflate button. This will deactivate the Max Inflate LED.
It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or
patient cleaning.
The Low Air Loss mode is only on the MA65 standard, bariatric and raised side rail system.