beautypg.com

Precision shunt calibration resistors, Recision, Hunt – Epson 9840 User Manual

Page 59: Alibration, Esistors

background image

I N T E R F AC E M O D E L 9 8 4 0 V E R 6 . 0 . 2

Precision Shunt Calibration Resistors

This diagram shows the connection of the precision internal shunt resistors.

The back panel switch (shown to the left of the two resistors) selects between 30 K

Ω

and 60 K

Ω.


The shunt check relay (shown to the right of the two resistors) connects the selected
resistor between the Excite High and the Cell Output High.

30 K

Ω

60 K

Ω

Cell Output High

Sense High

Excite High

Excite Low

Sense Low

Cell Output Low

LOAD
CELL














M O D E L 9 8 4 0 P G 5 8

P U B .

2 8 5 6 - 1 6