10. TECHNICAL SPECIFICATIONS
TECHNICAL FEATURES
Accuracy is indicated as [% of reading + number of digits]. It refers to the following
atmospheric conditions: a temperature of 23°C ± 5°C with a relative humidity < 75%.
Continuity
Test mode
Measuring range
(
Ω)
Resolution
(
Ω)
Accuracy
0.01 – 19.99
0.01
AUTO, R+, R-
20.0 – 99.9
0.1
±(2% Reading + 2 digit)
R+TIMER
R-TIMER
0.01 – 9.99
0.01
±(2% Reading + 2 digit)
Test current
>200mA DC up to 16
Ω (incluse the resistance of the calibration)
40mA DC from 16
Ω to 99.9Ω
Test current resolution
1mA
Open circuit Test Voltage
9V
Insulation resistance
Test voltage
(
Ω)
Measuring range
(M
Ω)
Resolution
(
Ω)
Accuracy
0.01 – 19.99
0.01
20.0 – 49.9
0.1
±(2% Reading + 2 digit)
50
50.0 – 99.9
0.1
±(5% Reading + 2 digit)
0.01 – 19.99
0.01
20.0 – 99.9
0.1
±(2% Reading + 2 digit)
100
100.0 – 199.9
0.1
±(5% Reading + 2 digit)
0.01 – 19.99
0.01
20.0 – 199.9
0.1
200 - 249
1
±(2% Reading + 2 digit)
250
250 - 499
1
±(5% Reading + 2 digit)
0.01 – 19.99
0.01
20.0 – 199.9
0.1
200 - 499
1
±(2% Reading + 2 digit)
500
500 - 999
1
±(5% Reading + 2 digit)
0.01 – 19.99
0.01
20.0 – 199.9
0.1
200 - 999
1
±(2% Reading + 2 digit)
1000
1000 - 1999
1
±(5% Reading + 2 digit)
Automatic selection of the measuring range
Open circuit voltage
1,1 x Rated voltage test
Short circuit current
<3,0mA - 500V
<2,0mA - 50V, 100V, 250V, 1000V
Rated measuring current
2,17 mA - 500V/230k
Ω
1mA - 1000V/1M
Ω
If the instrument measures a voltage higher than 30V, the instrument will not take a
measurement.
Safety standards
Insulation:
Class 2, double insulation
Over voltage category:
CAT III