Amprobe MEGATEST-1000 Megohmmeter User Manual
Page 12

5.1.2.1 "AUTO" MODE
CAL
LOW
Ω
1.07
Ω
219
mA
AUTO
At the end of the test, if the average
resistance value Ravg results to be
lower than 5
Ω the instrument emits a
double sound signal indicating the
positive outcome of the test and
displays one screen similar to the
screen alongside.
Average test
current value
Iavg.
Average
resistance
value Ravg.
SAVE
The test can be stored pressing the SAVE key twice (according to
paragraph 0).
5.1.2.2 "R+" or “R-" MODE
CAL LOW
Ω
1.07
Ω
219
mA
R+
R-
At the end of the test if a resistance
value R+ or
R- lower than 5
Ω was detected, the
instrument emits a double sound
signal indicating
the positive
outcome of the test and displays
one screen similar to the screen
alongside.
Resistance
value R+ or
R-.
Value of test
current I+ or I-.
Either the symbol R
+
or
the symbol R
-.
SAVE
The test can be stored pressing the SAVE key twice (according to
paragraph 0).
5.1.2.3 "R+TIMER" or “R-TIMER" MODE
CAL LOW
Ω
1.08
Ω
219
mA
5
s
R+
R-
TIMER
At the end of the test if a resistance
value R+Timer or R–Timer lower
than 5
Ω was detected, the instrument
(after the set time has elapsed) emits
a double sound signal indicating
the positive outcome of the test
displays one screen similar to the
screen alongside.
Duration time of
the test.
Value of the
test current I+
or I-.
The symbols R+ or R- are displayed.
SAVE The test can be stored by pressing the SAVE key twice (see paragraph 0).
11