Reset procedure, Default parameters, Default parameters low – Amprobe MEGATEST-1000 Megohmmeter User Manual
Page 25: Default parameters m, Memory default parameters
Example:
97 tests have been stored in the instrument.
If you want to cancel the tests from 43rd to 97
th
, you proceed as follows:
- Press RCL.
- Select with keys ▼, ▲ the memory location 43.
- Press CLR. You can see CLR blinking on the primary display, 43 on the
left secondary display and 97 on the righ secondary display.
- Press CLR. The tests from 43 to 97 has cancelled.
5.4 RESET
PROCEDURE
Reset of the INSTRUMENT and default parameters
Press DISP, CLR, RCL at the same time the rotary switch is turned on.
rES
The screen nearby is displayed
for 5 seconds, after which the
instrument emits a sound signal
and then displays the screen
relative to the function selected
with the rotary switch.
ATTENTION: the reset procedure will erase all the stored tests and it will set the default
parameters in the instrument.
Before conducting a reset procedure, download the stored tests on a PC.
5.5 DEFAULT
PARAMETERS
5.5.1 DEFAULT PARAMETERS LOW
Ω
Parameter
Reset default parameter
Mode AUTO
Calibration Offset
0 (CAL off)
Mode R+/R- TIMER
Timer 3s
5.5.2 DEFAULT PARAMETERS M
Ω
Parameter
Reset default parameter
Mode MAN
Test voltage
500V
Mode TIMER
Timer 60s
5.5.3 MEMORY DEFAULT PARAMETERS
Parameter
Reset default parameter
Parameter P
P = 1
Memory state
1(empty)