DM3068 Quick Guide

10

8. RS232 Port

Provides an access for multimeter controlling through PC software and outputs
P/F test result.

9. Power Fuse

The multimeter is already installed a power fuse before leaving factory. To
replace a new one, please:
1) Cut off the power supply.
2) Press down the block tongue using a straight screwdriver (in the direction of

the dotted arrow in figure below) and pull out of the fuse seat.

3) Select a proper voltage scale.
4) Replace a specified fuse.
5) Reinstall the fuse seat into the slot.

Figure 8 Replace the Power Fuse

WARNING

In order to avoid electric shock or fire, please just use the specified fuse
and make sure the fuse holder is in good connection and not shorted.

10. Voltage Selector

Select a proper voltage scale according to the used AC supply: 115 V or 230 V.

11. Power Switch

Connect or disconnect the AC supply. If the front power button is disabled (

 System  Cfg  Switch  OFF), turning on this switch will directly start up
the multimeter.