Cleaning / decontamination, Maintenance, Returns – Precision Medical Continuous Intermittent Vacuum Regulator User Manual
Page 9
![background image](https://www.manualsdir.com/files/858369/content/doc009.png)
8
CONTINUOUS INTERMITTENT
CLEANING / DECONTAMINATION
(As needed)
1. Attach a working Vacuum Regulator with a continuous regulated
mode to a minimum vacuum source of 15 inHg.
2. Mix cold disinfection / sterilization solution according to its
manufacturer’s directions.
3. Connect tubing as shown in Cleaning Illustration on previous page.
4. Turn the working Vacuum Regulator on to a continuous regulated mode.
5. Adjust the vacuum to a minimum of 120 mmHg.
6. Set the Vacuum Regulator to be cleaned to the “REG .” mode, and
set at 100 mmHg.
7. Allow cold disinfection / sterilization solution to pass through and collect
in Suction Canister. Procedure should continue for time recommended
by the manufacturer of the cold disinfection / sterilization solution for the
desired level of disinfection or sterilization.
8. Set working Vacuum Regulator to its maximum vacuum setting.
9. Thoroughly dry the internal components by drawing maximum
vacuum through the Regulator to be cleaned for at least 30 seconds
in “REG.” mode.
NOTE: If it is not possible to pass cold disinfection / sterilization
solution through the Regulator, then the passageways are
totally blocked and disassembly of the Regulator is required.
Be sure to follow your facilities’ Biohazard protocol.
MAINTENANCE
Before each use; visually inspect Vacuum Regulator for any sign of
damage, DO NOT USE if damaged.
RETURNS
Returned products require a Returned Goods Authorization (RGA) number,
contact Precision Medical, Inc. All returns must be packaged in sealed
containers to prevent damage. Precision Medical, Inc. will not be responsible
for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy
available on the Internet, www.precisionmedical.com.
Manuals available on our Website; www.precisionmedical.com