CONTINUOUS INTERMITTENT

CLEANING / DECONTAMINATION

(As needed)

1. Attach a working Vacuum Regulator with a continuous regulated

mode to a minimum vacuum source of 15 inHg.

2. Mix cold disinfection / sterilization solution according to its

manufacturer’s directions.

3. Connect tubing as shown in Cleaning Illustration on previous page.
4. Turn the working Vacuum Regulator on to a continuous regulated mode.
5. Adjust the vacuum to a minimum of 120 mmHg.
6. Set the Vacuum Regulator to be cleaned to the “REG .” mode, and

set at 100 mmHg.

7. Allow cold disinfection / sterilization solution to pass through and collect

in Suction Canister. Procedure should continue for time recommended

by the manufacturer of the cold disinfection / sterilization solution for the

desired level of disinfection or sterilization.

8. Set working Vacuum Regulator to its maximum vacuum setting.
9. Thoroughly dry the internal components by drawing maximum

vacuum through the Regulator to be cleaned for at least 30 seconds

in “REG.” mode.

NOTE: If it is not possible to pass cold disinfection / sterilization

solution through the Regulator, then the passageways are

totally blocked and disassembly of the Regulator is required.

Be sure to follow your facilities’ Biohazard protocol.

MAINTENANCE

Before each use; visually inspect Vacuum Regulator for any sign of

damage, DO NOT USE if damaged.

RETURNS

Returned products require a Returned Goods Authorization (RGA) number,

contact Precision Medical, Inc. All returns must be packaged in sealed

containers to prevent damage. Precision Medical, Inc. will not be responsible

for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy

available on the Internet, www.precisionmedical.com.