4

CONTINUOUS INTERMITTENT

OPERATING INSTRUCTIONS

CAUTION

Inspect the Vacuum Regulator for visual damage before use,

DO NOT USE if damaged.

NOTE: • Overflow protection should be used with the Vacuum Regulator.

(i.e. Filter, Vac Trap, Canister equipped with float shutoff).

• Gauges operate independently; if the digital gauge fails, the

analog gauge will still function.

1. Turn the Selector Knob to the “OFF” position.

2. Attach the Vacuum Regulator to the vacuum source.

A. REG MODE (Regulated Mode)

ALL MODELS

1. Turn the Selector Knob to the “REG.” position.

2. Block the bottom port of the Regulator.

3. Using the Regulator Knob, set the desired vacuum.

To

INCREASE vacuum - Turn Knob CLOCKWISE

To

DECREASE vacuum - Turn Knob COUNTERCLOCKWISE

B. INT. MODE (Vacuum cycles ON and OFF.)

1. Turn the Selector Knob to the “REG.” position, to select

desired vacuum level.

2. Turn the Selector Knob to the “INT.” position.

NOTE: Intermittent cycles starts in the OFF phase, therefore a

delay occurs before the intermittent cycle begins.

3. Turn the Selector Knob to the “OFF” position to turn the Regulator off.

WARNING

• ALWAYS confirm vacuum setting prior to performing procedure.

• When turning the Vacuum Regulator to “REG.” or “INT.” from any

position, the vacuum level will return to its previously regulated setting.

Full Line Vacuum is present between settings.

• Vacuum levels will remain the same when switching from one

mode to the other.

CAUTION

DO NOT operate the Vacuum Regulator when the collection

canister is “full”. This may cause loss of vacuum and damage to

the Vacuum Regulator. This will

void the warranty.