Specifications – Precision Medical Continuous Intermittent Vacuum Regulator User Manual
Page 4
3
CONTINUOUS INTERMITTENT
SPECIFICATIONS
GAUGE RANGE:
P
M3300:
0 - 200 mmHg - Full Vacuum
*PM3300E:
0 - 200 mmHg (0 - 26 kPa)
*PM3300EHV:
0 - 300 mmHg (0 - 40 kPa)
PM3300HV:
0 - 300 mmHg - Max Vacuum
PM3400:
0 - 150 mmHg
*PM3400E:
0 - 150 mmHg (0 - 20 kPa)
*Counterclockwise direction
GAUGE ACCURACY:
Analog:
± 5% of MAX
Digital/Analog, Dual Gauge:
Digital Display:
± 1% of Full Scale
Analog Gauge:
± 5% of MAX within ref. Indicator
VACUUM PORTS:
1/8 NPT Female
MODES:
REG. - (Regulated) provides an adjustable,
continuous vacuum level
OFF - No Vacuum
INT. - (Intermittent) provides an adjustable vacuum
level that cycles between ON and OFF
FLOW:
Models
Mode
Max Flow
PM3300:
REG
51 l/min
PM3400:
50 l/min
MAXIMUM FLOW
IS OBTAINED WITH A VACUUM SOURCE OF 21” Hg (71.1 kPa)
MAXIMUM VACUUM:
PM3300: REG. Mode @ Full Vac-396 mmHg (53 kPa)
PM3300HV: REG. Mode @ Max Vac-396 mmHg (53 kPa)
PM3400: Restricted to 170 mmHg ± 10 mmHg (1.3 kPa)
INTERMITTENT CYCLE TIME:
Factory set at sixteen (16) seconds ON
and eight (8) seconds OFF ( Reference only)
Operating Environmental Limits:
0°F to 122°F (-18°C to 50°C)
Recommended Environmental Operating Limits: 55°F to 85°F (13°C to 29°C)
Storage Environmental Limits:
Temperature Range:
-4°F to 140°F (-20°C to 60°C)
Humidity:
Max 95% Noncondensing
Battery: 3 Volt Lithium, AA (Digital Vacuum Gauge Models ONLY)
Specifications are subject to change without prior notice.