Caution, Vacuum regulator cleaning illustration – Precision Medical Continuous Intermittent Vacuum Regulator User Manual

Page 8

background image

7

CONTINUOUS INTERMITTENT

CAUTION

DO NOT autoclave or immerse in liquid.

This will cause damage to the Vacuum Regulator and will

void the

warranty.

DO NOT set the Vacuum Regulator to the “INT“(Intermittent) mode when

cleaning. Pulling cleaning fluids thru the Intermittent mode will damage the

timing module and void the warranty.

If Vacuum Regulator becomes internally contaminated, warranty is voided,

DO NOT send back to Precision Medical, Inc. for repair. Follow your facilities

contaminated equipment protocol.

This Vacuum Regulator contains magnetic, ferrous material that may affect

the results of an MRI.

Be sure all connections are tight and leak free.

VACUUM REGULATOR CLEANING ILLUSTRATION

(Cleaning/Decontamination Instructions on

page 8)

Vacuum Regulator

To Be Cleaned

Working

Vacuum

Regulator

Suction

Tubing

Vacuum

Port

Patient

Port

Suction

Collection

Canister

Cold

Disinfection/

Sterilization

Solution

Suction

Tubing