Caution, Vacuum regulator cleaning illustration – Precision Medical Continuous Intermittent Vacuum Regulator User Manual
Page 8
![background image](https://www.manualsdir.com/files/858369/content/doc008.png)
7
CONTINUOUS INTERMITTENT
CAUTION
•
DO NOT autoclave or immerse in liquid.
This will cause damage to the Vacuum Regulator and will
void the
warranty.
•
DO NOT set the Vacuum Regulator to the “INT“(Intermittent) mode when
cleaning. Pulling cleaning fluids thru the Intermittent mode will damage the
timing module and void the warranty.
•
If Vacuum Regulator becomes internally contaminated, warranty is voided,
DO NOT send back to Precision Medical, Inc. for repair. Follow your facilities
contaminated equipment protocol.
•
This Vacuum Regulator contains magnetic, ferrous material that may affect
the results of an MRI.
•
Be sure all connections are tight and leak free.
VACUUM REGULATOR CLEANING ILLUSTRATION
(Cleaning/Decontamination Instructions on
page 8)
Vacuum Regulator
To Be Cleaned
Working
Vacuum
Regulator
Suction
Tubing
Vacuum
Port
Patient
Port
Suction
Collection
Canister
Cold
Disinfection/
Sterilization
Solution
Suction
Tubing