beautypg.com

Energizer PORTABLE GAS DETECTION INSTRUMENT User Manual

Page 45

background image

45

CHLORINE DIOXIDE SPECIFICATION

Range:

0.00 to 2.00 ppm

Response Time:

T80 < 60 seconds

Operating Temperature Range:

-20

o

to +50

o

C

(-4

o

to +122

o

F)

Graphic Display Temperature Range:

-10 to +60

o

C

(+14 to +140

o

F)

Operating Humidity:

15 to 90% RH, Non-Condensing

Recommended Storage Temperature:

0 to 20

o

C

(32 to 68

o

F)

Sensor Accuracy:

±3% of Applied Gas

Linearity:

±3% of Full Scale

Battery Life:

About 100 Hours - Non-Alarming
About 800 Hours in Battery Save Mode

Sensor Cross Sensitivity:

Chlorine at 3 ppm: 1 ppm
Carbon Monoxide at 300 ppm: 0 ppm
Hydrogen Sulfide, 15 ppm: -3 to 0 ppm
Sulphur Dioxide, 5 ppm: 0 ppm
Nitric Oxide, 35 ppm: 0 ppm

Consult Factory for additional information.