beautypg.com

Energizer PORTABLE GAS DETECTION INSTRUMENT User Manual

Page 27

background image

27

the “OK” pushbutton to return to the Operating Mode.
To change the setpoint, depress the “CHANGE” pushbutton. One of the
following screens will be displayed:

ALARM SETPOINT

For Toxic Gases

HI ALARM = <_ >ppm
UP DOWN

CHANGE O

2

ALARM

For O

2

HI ALARM = 23.5%

LO ALARM = 19.5%

LO HI

For toxic gases, choose “UP” or “DOWN” to change the corresponding
“ALARM SETPOINT”. For example, choose “UP” and one of the following
screens will appear:

ALARM SETPOINT

For Toxic Gases

HI ALARM = <_ >ppm
UP SET

Choose “DOWN” and one of the following screens will appear:

ALARM SETPOINT

For Toxic Gases

HI ALARM = <_ >ppm
DOWN SET

The ppm value will flash while either of these screens is displayed. Each
time the “UP” or “DOWN” pushbutton is depressed, the setpoint will be
changed by 1 ppm. Holding the pushbutton automatically repeats the action.
For oxygen, choose “LO” or “HI” to change the corresponding “ALARM
SETPOINT”. For example, choose “LO” and the following screen will ap-
pear: