Energizer PORTABLE GAS DETECTION INSTRUMENT User Manual
Page 10
10
5.0 SWITCH FUNCTIONS AND DISPLAYS
All operations on the MINI-SA Portable Detector are controlled by two push-
buttons located on the bottom of the unit. The operations vary depending on
the operating mode and the graphic display appearing above the two switches.
The unit sounds a brief audible “beep” whenever either pushbutton is de-
pressed while the instrument is on. Whenever the instrument is turned ON or
OFF, three successive audible tones will sound and the Red Alarm LED will
flash. The Green Status LED will flash once every fifteen (15) seconds during
instrument operation. When in the BATTERY SAVER mode, the Green Sta-
tus LED flash rate increases to once every five (5) seconds.
When the unit is in the operating mode, depressing the Display Backlight
Pushbutton will illuminate the graphic display for fifteen (15) seconds. A light
bulb icon for the Backlight is located above the pushbutton on the operat-
ing mode screen. Depressing the button again extinguishes the Backlight.
The Backlight is sustained ON or OFF while the MENU operations are in
progress, except for calibration. In an alarm condition, the Backlight flashes
for the duration of the alarm indication.
When in the BATTERY SAVER mode, the instrument graphic display will be
blank. Depressing either pushbutton will turn on the graphic display. The
display remains on for thirty (30) seconds after the last pushbutton opera-
tion.
WARNING
USE OF THE DISPLAY BACKLIGHT FOR EXTENDED PERIODS OF TIME
WILL REDUCE BATTERY SERVICE LIFE.
COMPONENT LOCATIONS
FIGURE 3
PUSHBUTTONS (2)
LED
(GREEN OR RED
DEPENDING
SENSOR
GRAPHIC