Maintenance

Re-fitting MiniPID pellet and lamp

Caution! Never refit a damaged lamp

1.

Place the lamp inside the O-ring seal in the pellet as illustrated. Twisting the lamp slightly during
insertion will help to ensure the lamp window is snug against the pellet’s front electrode. The lamp
should be freely supported by the O-ring.

2.

Lay the pellet front face down on a clean, flat surface and then screw the lamp down into the O-ring
until it firmly abuts against the front electrode face

– this is most important. Then bring the MiniPID

body carefully down over the lamp so as not to disturb its positioning within the pellet and then push
the body firmly onto the face down pellet so that it clicks into place.

3.

Refit the sensor into the sensing equipment.

4.

Re-

calibrate the equipment in accordance with manufacturer’s instructions.