beautypg.com

Gas detection specifications (specific), Specification – Ion Science MiniPID User Manual

Page 19

background image

MINIPID 3PIN MANUAL

Ion Science Ltd

Page 19 of 27

Unrivalled Detection. www.ionscience.com

0

20

40

60

80

100

120

10

100

1000

10000

liniear

it

y

,

10

0

p

p

m =

10

0%

isobutylene, ppm

Specification

For optimal performance, Ion Science recommends an operational voltage of 3.3V.

Gas Detection Specifications (specific):
Transducer Details
(ref Isobutylene)

MiniPID PPB

MiniPID PPM

Minimum Detection Level

S

D

1 ppb

100 ppb

Linear Range (±3% deviation)

R

L

full range

100 ppm

Minimum over-range

R

O

50 ppm

6,000 ppm

Over range typical

R

T

80 ppm

10,000 ppm

Sensitivity (Linear range)

S

> 25 mV/ppm

> 0.6 mV/ppm

Full Stabilisation Time (PPB 20 ppb, PPM 100 ppb)

T

S

5 min

10 s

Warm-up Time

T

W

< 5 s

< 5 s

Response Time in diffusion mode(t

90

)

T

R

< 3 s

< 3 s

Offset Voltage

V

OS

60-70 mV

50-54 mV

Gas Detection Specifications (general):
Target Gases

VOC’s with ionisation potentials < 10.6 eV

Temperature related response variance

±10% between -20 and 60

o

C, vs 20

o

C response

(please check item 4 on page 14)

Relative humidity range

0 to 99% RH, non-condensing

Product Specifications (general):
Lamp replacement

User replaceable (10.6 eV)

Electrode Pellet

User replaceable

Onboard filter (within disposable pellet)

Removes liquids and particulates

Package Type

Alphasense

TM

CH-A3,City Technology

TM

4P,

20 mm dia x 16.6 mm high

Weight

< 9 g

Positional Sensitivity

None

Warranty

12 months from date of shipment.

(Please see page 25 for details on extended warranty)

Typical MiniPID Linearity