Ion Science Tiger Select User Manual
Page 9

TIGER MANUAL
Ion Science Ltd
 
 
 
 
 Page 9 of 20 
Unrivalled Detection. www.ionscience.com
Fitting the tube holder
 
Always ensure the filter tube is visible while viewing the display screen. If necessary the filter tube assembly 
should be removed and refitted. 
 
1. Unscrew and remove the Filter Cap. 
 
 
 
 
 
 
 
 
 
2. Remove the filter clamp and tube holder together. 
 
 
 
 
 
 
 
 
 
 
3. P
lace the O’ring on the filter lamp and push the
assembly in to the filter housing with the window facing 
forward. Continual pressure may be required to keep the 
assembly in place while tightening the filter cap. 
 
 
 
 
 
 
 
The filter disk should be replaced if it appears dirty, or is disturbed from its seating position. 
The filter disks can be fitted either way round however the orientation should never be reversed once used. 
 
 
 
 
 
 
 
 
 