TAC mode routine
The TAC soft key is used to ENTER and EXIT from TAC mode.
1. To select TAC mode press the TAC soft key then press then press the
ENTER key to continue.
2. A tick /
check ’’ will appear to confirm the selection
3. Press the Esc key to clear the confirmation
4. The TAC response factor will be displayed temporarily before entering
TAC mode operation.
5. TAC mode automatically selects a specific instrument setup, to ensure
best results. The TAC response factor, high sensitivity and data logging
functions become available.
This mode offers single point data logging, multiple data logging, STEL and Tube mode.
Important
The TAC gas used within TAC mode has a STEL set to 1 ppm, this level has been chosen based the low
STEL levels often associated with Benzene gas. The Tiger Selects TAC STEL however is not supported by
nationally recognised bodies who publish official levels
6. To exit TAC mode Press the TAC soft key and then ENTER.
7. A tick /
check ’’ will appear to confirm the selection, then press the Esc
key.
8. The gas repose factor will be displayed temporarily before exiting TAC
mode.