Working with scaling factors, Error messages – HEIDENHAIN ND 510 v.2 User Manual
Page 5
Working with Scaling Factors
The ND 510 and the ND 550 can display the axis traverse lengthened or shortened by
a scaling factor. You enter a scaling factor separately for each axis in the user
parameter P12, then activate the scaling factor function with the user parameter
P11. SCL is highlighted.
Error Messages
Message
Cause and effect
Traverse distance with datum setting function (SPEC FCT) is
too short
Incorrect input value
Input frequency too high for encoder input
(will occur for example when traverse speed too high)
Internal counter overflow
Error while crossing over reference marks
To clear the error message: Switch of the display unit.
Should any of these error codes recur, contact your
HEIDENHAIN service agency.
Compensation values for nonlinear axis error compensation have
been erased
Datums have been erased
Erase the operating parameters
If all decimal points light up, the measured value is too large or too small.
Set a new datum.
To clear error message
:
When you have removed the cause of the error,
➤ press CL.