Deployment, Troubleshooting – Grass Valley SHED Jan 16 2015 User Manual

Page 20

background image

16

SHED and HDX Features

Deployment

Deployment

The figure shows a typical deployment of a SHED-C and SHED-BS. In this case, the SHED-C is
in passive mode, as a local power supply is connected directly to the camera.

Fig. 3-7: SHED-C and SHED-BS Deployment Diagram

Troubleshooting

Symptom

Possible Cause

Corrective Action

No AC IN indication

Power source

Verify correct VAC selector
setting
Be sure power switch is On:

1 = On; 0 = Off

Check fuses

Camera will not power.
RED LEDs for Cable Short
or Cable Open

Bad SMPTE Hybrid cable

Check for electrical problems
with hybrid cable and/or
replace the hybrid cable that
connects the CCU to the SHED-
BS or the CAMERA to the the
HDX/SHED-C

No Optical Link. Optical
Power Meter reads
>20bDm

Bad fiber link

• Verify CCU/CAM is on and

optics are working

• Check all fiber to ensure that

connectors are clean and that
there are no bend-radius
issues along the run. Minimize
in-line connectors and
patches, if possible.

HV not being applied (No
HV LED = GREEN)

Camera system not
powered. Improper
camera system

Make sure that the camera
system is supported

Make sure that the camera
system is powered

CAM stays powered ON
when CCU is powered off

Switch in wrong position
at HDX

If you wish the camera to power
down when the CCU is turned
(only if this option is installed)