Example images, Pyramid probes: new core inspection – Cascade Microtech Pyramid Probes User Manual
Page 9

PN 153-690-3
www.cascademicrotech.com
Pyramid Probes: New Core Inspection • 9
Example Images
Probe Tips
(cont.)
Do the probe tips match the Outgoing Report Tip Measurement Data specs for height and
diameter dimensions?
For new cores, refer to the specs in the Pyramid
Probe
Outgoing Report Tip Measurement Data
for the core you are inspecting.
Do adjacent probe tips have a height difference greater than 2μm?
Adjacent probe tips with height difference greater than 2μm are considered a reject on new cores, unless by
design.
Low Magnification Scope
High Magnification Scope
Entire Probe
Core Window
Entire Membrane
Beams @ 50x
Probe Face
Wing North and East
Wing South and West
Probe Tips@ 50x
Adjacent tip 1.5 μm
shorter, acceptable
Adjacent tip 2.5 μm
shorter, not acceptable
Pyramid Probes: New Core Inspection