Setup, Amount measured by device, Actually delivered quantity – ARAG DIGIWOLF - Battery powered version User Manual
Page 11: Bd a, X [constant indicated on flowmeter body
11
4.1.2
Manual calibration
To set the flowrate constant manually, calculate and set the correct constant using the following formula:
[amount measured by device]
[actually delivered quantity]
Calibration
Man. calibration
Auto calibration
600
1) Access Manual calibration menu (Setup menu > Calibration > Man. calibration).
In Calibration menu, selecting Manual calibration, under the item the curren-
tly set constant will be displayed.
Press
OK
to access value modification.
Flowrate constant
Min value
Max value
C
1
50000
B
D
A
0600
0
2) Using the suitable keys, set the value of the flowmeter constant:
A)
Press to toggle from one digit to another
B)
Press in succession to modify the value of the digit highlighted by the cursor
C)
Press to save changes or
D)
Press to exit current page without confirming
changes. Refer to the label on the body (Fig. 21).
The flowmeter body must be assembled with the arrow on the label
facing the flow direction.
CODE
Approximate constant
4628405
250
4628506
132
4628707
64
APPROXIMATE CONSTANT
XXX
APPRO
XIMAT
E CON
STANT
XXX
Fig. 19
Fig. 20
Fig. 21
x [constant indicated on flowmeter body]
Tab. 3
SETUP