Function 80 – Reset memory
This function will reset the memory with the exception of the part used for saving programs.
The PF6 will enter stand-by mode, and when switched on again the built-in parameters will be valid.
IF THE PF6 DOES NOT WORK OR DOES NOT OPERATE AS EXPECTED, ACTIVATE FUNCTION
80.
7 Print-outs
PF6 can be connected to a printer (via RS-232) which can produce three different types of print-outs.
7.1 Print current parameters
Function 29 prints the current parameters, and print-outs should be made after programming and
calibrating PF6 and resetting function 8.
The print-out is started in the following way: <2>+<9>+
Press the
7.2 Print current status
Function 24 prints the current status, and print-outs should be made after completion of the filling
series.
The printing is to be started in the following way: <2>+<4>+
Press the
After the status print-out the printer memory is deleted automatically.
7.3 Print programs
Up to five programs can be stored in PF6, and a list of saved programs can be printed by means of
function 34.
The printing is to be started in the following way: <3>+<4>+
Press the