Function 1 - volume, Function 2 - tube diameter, Function 3 - velocity – Watson-Marlow PF6 User Manual
Page 18: Function 4 - acceleration/deceleration, Instruction handbook

INSTRUCTION HANDBOOK
PF6
PF6 IH EN 74-105-022 v1.04.doc
Version: 1.04
Page 18 of 30
Function 1 - Volume
Value: ml
Function 1 informs the system of the volume to be filled.
Value
Option
Min
Max
0.01
9999.9
ml. or gram
The entered value must be between 0.01 and 9999.9.
Function 2 - Tube diameter
Value: Inside diameter of the tube in mm.
Drive type
Tube inside diameter in mm
PF6
0.5
0.8
1.2
1.6
3.2
4.8
6.0
8.0
Function 3 - Velocity
Value: Revolutions per minute (rpm).
Velocity range depends on tube size applied.
Range:
Tube Sizes
Max. Velocity
Max. acceleration
0.5 - 0.8 – 1.2 – 1.6
400
100
3.2
400
100
4.8 - 6.0 – 8.0
400
100
The fastest filling will be carried out at the highest velocity setting but the velocity should always be
adjusted to suit the characteristics of the product and to reduce splashing or foaming.
Function 4 - Acceleration/deceleration
Value: An integer number.
This function offers a choice of values between 1 and 100 dependent on the tube size and drive;
1 = slowest, 100 = fastest.
Tube Sizes
Max. Velocity
Max. acceleration
0.5 - 0.8 – 1.2 – 1.6
400
100
3.2
400
100
4.8 - 6.0 – 8.0
400
100