23.4 313D and 314D flow rates

Flow rates were obtained using silicone tubing with the pumphead rotating clockwise,
pumping water at 20°C with zero suction and delivery pressures. For critical applica-
tions determine flow rates under operating conditions.

Flow rates, 313D, 1.6mm wall (ml/min)

bore

mm

0.5

0.8

1.6

3.2

4.8

6.4

8.0

in

1/50

1/32

1/16

1/8

3/16

1/14

5/16

#

112

13

14

16

25

17

18

15-400 rpm
3-400 rpm
1.5-220 rpm

0.45-12
0.09-12

0.05-6.6

1.1-28

0.21-28
0.11-15

4.1-110

0.81-110

0.41-59

15-400

3.0-400
1.5-220

33-880

6.6-880
3.3-480

54-1400
11-1400

5.4-790

75-2000
15-2000

7.5-1100

Flow rates, 314D, 1.6mm wall (ml/min)

bore

mm

0.5

0.8

1.6

3.2

4.8

6.4

8.0

in

1/50

1/32

1/16

1/8

3/16

1/14

5/16

#

112

13

14

16

25

17

18

15-400 rpm
3-400 rpm
1.5-220 rpm

0.45-12
0.09-12

0.05-6.6

0.90-24
0.18-24
0.09-13

3.8-100

0.75-100

0.38-55

13-340

2.6-340
1.3-190

29-760

5.7-760
2.9-420

45-1200

9.0-1200

4.5-660

60-1600
12-1600

6.0-880

Note: 314D pumpheads should run no faster than 300 rpm when in continuous use.