27

RS232 settings

Baud

9600

Stop bits

2

Data bits

8

Parity

None

Flow control

None

Echo

On

Pump inteface
pin

Function

1

-

2

RX (Receive data)

3

TX (Transmit
data)

4

-

5

GND (Ground)

6

-

7

-

8

Reserved

9

-

RS232 command modes

These are the codes to control the pump with the RS232 serial link. They must be sent
to the pump from a computer serial port (or equivalent).

Com-
mand

Function

Com-
mand

Function

1SPxxx

Set pump speed to xxx

1RC

Reverse direction

1SI

Increase speed by 1rpm

1RR

Set clockwise direction

1SD

Reduce speed by 1rpm

1RL

Set counter-clockwise
direction

1GO

Start pump

1RS

Show all pump information

1ST

Stop pump

1ZY

Show running status of the
pump. STARTed 1 or
STOPped 0

Always terminate each command with a RETURN (ASCII CHR13).

Notes on control codes

There should be at least 10mS between commands.

The command RS to show all pump information will return the following text string:

[pump type] [speed] [CW / CCW rotation] [stopped/running, 0 /1] [! delimiter]
e.g. 323Du 110 CW 1 !

Note: both 9 and 25-way D shells are earthed.