Watson-Marlow 505Dz User Manual
Page 9

9
DOSE
Pressing Enter at DOSE while in the Main menu will return the message:
5 . 0 0 M L
E N T E R
D O S E
Enter the required dose using the key pad numbers.
Any dose from 0.001ML to 9999ML can be set. Up to three decimal places are
allowed, limited to a maximum of 4 characters.
On Enter, display changes to:
_
2 . 5
S E C O N D S
E N T E R
I N T E R V A L
Enter the required interval using the key pad numbers.
Enter enters time up to 3445 seconds, with up to one decimal place. The
minimum time is to be 0.1 seconds.
Set to 0 SECONDS to start a single dose with Start or an external switch.
The number of doses will be requested only if a time interval has been entered:
_
2 5 D O S E S
E N T E R
D O S E S
Enter the required number of doses using the key pad numbers
Enter enters number up to 9999.
On Enter the display changes to:
2 1 5 R P M
C C W
P R E S S
E N T E R
To change the speed enter the required rpm using the numbers on the keypad or
press /\ or \/
/\ / Max sets maximum RPM.
\/ /Max sets minimum RPM.
CW/CCW can be changed by pressing CW/CCW.
Changing the RPM must be confirmed by pressing Enter, after which the RPM
will be displayed.
Pressing Enter calls up confirmation of your selection then requests press Start.
3 5 M L
1 6 0 D O S E
C C W
2 1 5 R P M
2 . 5
S E C S
I N T E R V A L
P R E S S
S T A R T
Entering new
program
Single dose
command
Start the dose