CALIBRATE

In the main menu, Step to CALIBRATE, then press Enter.
Depending upon whether the last calibration was a calibration dose setting in
ml/min or by specifying the pumphead and tubing fitted one of the following
screens will be shown:

5 0 1 R L

8 . 0 M M

R E C A L I B R A T E

or

5 0 1 R L

9 . 8 M L / R V

R E C A L I B R A T E

In both cases press Enter, and the screen will request

C A L I B R A T I O N

D O S E

T U B E

S I Z E

Step to the option required then press Enter.

When running a calibration dose it should be run in the duty configuration and
with the duty fluid. The program will run the pump for up to four minutes or you can
Stop it sooner, however the longer you allow it to run the better. The minimum run
time should be 15 seconds or 10 revolutions of the rotor. You can enter the value
of the calibration run in ml. On completion of the procedure the screen will
confirm the calibration figures and will also confirm the selection as well as
displaying the maximum allowable rpm.

5 0 1 R L

9 . 9 9 9 M L / R V

M A X I M U M

3 0 0 R P M

Calibration is now complete, and the display will return to the main menu.

SETUP

At the Main menu, Step to SETUP, press Enter and a choice will be offered.
Step to the required menu item and press Enter.

You are able to turn the audible signal off using this menu item. The default
condition is on.

Adjusting the ramp value will alter the acceleration/deceleration of the rotor.
There is a separate selection for the start of a dose and at the end of a dose. A
ramp setting of 2 is the factory default where there is no delay of acceleration,
the maximum setting is 5. To select the ramp enter the value on the keypad and
press Enter. The display will confirm your selection for three seconds.

A gradual start may be useful for applications dispensing sensitive fluids, or to
avoid splashing. A gradual stop can often decrease the dripping after the pump
has stopped.

By adjusting the drip function you can control a reverse rotor direction at the end
of a dose. A 0 value has no reverse, the maximum reverse is 5. To select the
drip enter the value on the keypad. After the selection has been made and
Entered the display will confirm your selection.

This provision is to select the language shown on the display.

By following the prompts, all settings can be restored to the factory defaults.

Calibration dose

Beep

Ramp

Drip

Language

Default