beautypg.com

Watson-Marlow 900 Series User Manual

Page 12

background image

Page

12

Continue to rotate the rotor until the

lower tube connector is flush with

the port outer face. Tighten the

connector securing nuts to avoid

pulling the tube back through the

lower port.
Repeat connector fitting procedure

for the other end of the tube in the

upper casing port. Rotate the rotor

to ensure that the tube is seated in

the pumphead correctly.

Swing the pumphead door closed

and lock shut using the door

locking screws and the Allen key

supplied. Locate and secure the

lubricant drainage plug.


Replace the lubricant drainage plug
then fill the pumphead with lubricant
to the specified volume.

The window in the pumphead door
is a motion window and provides a
visual indication of the rotation of
the rotor within the pumphead. This
window should not be used as
lubricant level indicator.

Locate and secure the lubricant filler
plug.