beautypg.com

2 common practice commands, Burst mode, Operating modes – Tecfluid MX4B User Manual

Page 22: Write protection

background image

22

Number Command

33

Read Device Variables

34

Write Primary Variable Range Values

35

Write Primary variable Damping Values

38

Reset Configuration Changed flag

40

Enter/Exit Fixed Current Mode

44

Write Primary Variable Units

48

Read additional device status

5.2

Common practice commands

Command implemented are the following:

6 BURST

MODE

MX4H converter does not support Burst Mode.

7 OPERATING

MODES

The converter MX4H implements fixed current mode.
To change to this mode Command 40 can be used.
The output current can be set to any fixed value between 4 mA and 20 mA.
By sending Command 40 to the MX4H with a value of “0” the fixed current mode will be
exited.
This mode is automatically exited on a device reset or power up.
Current signalling can be disabled using Command 6. In this case the current output will
be set to 4 mA.
Current signalling is normally disabled when the converter is part of a network coexisting
with other instruments.
When current signalling is disabled using the “Loop Current Mode = 0” in Command 6, the
only way in which current signalling can be enabled is through the use of Command 6 and
setting “Loop Current Mode = 1”.
If there is a power up, current signalling keeps this operating mode.

8 WRITE

PROTECTION

Write protection is provided, selected by an internal jumper.

When the jumper is present, all commands are accepted. When the jumper is absent, no
write commands or local configuration changes are accepted.

This manual is related to the following products: