beautypg.com

Command summary – Sensoray 7429 User Manual

Page 41

background image

COMMAND SUMMARY

DEFINE CHANNEL SENSOR - declare sensor type connected to a channel

Custom resistive sensors only:

command:

(16 + CHAN),(SENSOR DEFINITION
CODE),
(HIGH MSB),(HIGH LSB),
(MID MSB),(MID LSB),
(LOW MSB),(LOW LSB)

response:

NONE

Strain/pressure gages only:

command:

(16 + CHAN),(SENSOR DEFINITION
CODE),
(V HIGH),(V LOW),
(P HIGH),(P LOW),
(R HIGH),(R LOW)

response:

NONE

All other sensors:

command:

(16 + CHAN),(SENSOR DEFINITION
CODE)

response:

NONE

READ CHANNEL DATA - read linearized sensor data from a single channel

command:

(CHAN)

response:

(HIGH DATA),(LOW DATA)

SET CHANNEL ALARM LIMITS - specify sensor limits for a single channel

command:

(32 + CHAN),
(HIGH LIMIT MSB),(HIGH LIMIT LSB),
(LOW LIMIT MSB),(LOW LIMIT LSB)

response:

NONE

READ ALARM FLAGS - read alarm status for all channels

Channels 0-7:

command:

(48)

response:

(HIGH ALARMS),(LOW ALARMS)

Channels 8-15:

command:

(49)

response:

(HIGH ALARMS ),(LOW ALARMS)