beautypg.com

2 meter – OWON HDS1021M User Manual

Page 61

background image

OWON HDS1021M Handheld DSO & DMM 10-Appendix

Probe

1X position

10X position

Bandwidth

Up to 6 MHz (DC)

Up to full bandwidth (DC)

Attenuation rate

1: 1

10: 1

Compensation range

15pf~35pf

Input resistance

1MΩ±2% 10MΩ±2%

Input impendence

85pf~115pf 14.5pf~17.5pf

Input voltage

200 V DC+Peak AC

600 V DC+Peak AC

10.1.2 Meter

Voltage (VDC)

Input Impedance: 10MΩ.

Max. Input Voltage: 1000V (DC or AC peak-to-peak value)

Range Accuracy

Resolution

400.0mv 100uV

4.000V 1mV
40.00V 10mV
400.0V 100mV

1000.0V

±

1%±2 digit

1V

Voltage (VAC)

Input Impedance: 10MΩ.

Max. Input Voltage: 750V(AC, virtual value)

Frequency range: from 40Hz to 400Hz.

Display: Virtual value of the sine wave

Range Accuracy

Resolution

4.000V 1mV
40.00V 10mV
400.0V

±

1%±3 digits

100mV

750.0V

±

1.5%±3 digits

1V

Direct Current (DC)

Range Accuracy

Resolution

40.00mA

±

1%±1 digit

10uA

400.0mA

±

1.5%±1 digit

100uA

20A

±3%±3

digits

10mA

54