ONICON F-3200 Series User Manual
Page 15
11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax (727) 442-5699 • [email protected]
F-3200 Flow Meter Manual 05/14 - 0671-12 / 18325
Page 15
3.2 MECHANICAL INSTALLATION
IMPORTANT NOTE
F-3200 transmitters and sensor bodies are two parts of one uniquely calibrated system and must
be installed together. Mixing components from other systems will result in significant calibration
errors.
i
3.2.1 Standard Transmitter Dimensions