Specifications

System throughput

Table 2. System throughput specifications

Condition Calibration

coefficients

Max

ADC

rate

1. Single channel, single input range.

Per specified range

2.0 MS/s

2. Multiple-channel, single-input range: ±5 V,

±2.5 V, ±1.25 V, 0 to 5 V, 0 to 2.5 V

Per specified range

1.5 MS/s

3. Multiple channel, single input range: ±0.625 V, 0

to 1.25 V

Per specified range

750 kS/s

4. Single channel, multiple input ranges.

Default to value for cbAInScan() range

500 kS/s

5. Multiple channels, multiple ranges. All samples in

unipolar or bipolar bode.

Default to value for cbAInScan() range

500 kS/s

6. Multiple channels, multiple ranges. All samples in

unipolar and/or bipolar bode.

Default to value for cbAInScan() range

500 kS/s

7. Multiple-channel, switching unipolar/bipolar

mode, single-input range.

Default to value for cbAInScan() range

750 kS/s

Note 1:

For conditions 1-3, specified accuracy is maintained at rated throughput. Conditions 4-7 apply
calibration coefficients which correspond to the range value selected in

cbAInScan()

. These

coefficients remain unchanged throughout the scan. Errors of up to 25 counts may be incurred
when switching gains while in bipolar or unipolar mode only (conditions 4 and 5). Errors of up to
500 counts may be incurred when mixing unipolar/bipolar modes (conditions 6 and 7).

Accuracy

A 100 kS/s sampling rate, single-channel operation and a 60-minute warm-up. Accuracies are listed for
operational temperatures within ±2ºC of internal calibration temperature. Calibrator test source high side tied to
channel 0 and low side tied to low-level ground at the user connector

Table 3. Analog input — absolute accuracy specifications

Table 3

Range Absolute

accuracy

±5.000 V

±6.0 LSB

±2.500 V

±8.0 LSB

±1.250 V

±8.0 LSB

±0.625 V

±10.0 LSB

0 V to +5.000 V

±8.0 LSB

0 V to +2.500 V

±11.0 LSB

0 V to +1.250 V

±13.0 LSB

Table 4. Analog input — typical accuracy specifications

Range Typical

accuracy

±5.000 V

±5.5 LSB

±2.500 V

±7.5 LSB

±1.250 V

±7.5 LSB

±0.625 V

±9.5 LSB

0 V to +5.000 V

±7.5 LSB

0 V to +2.500 V

±10.5 LSB

0 V to +1.250 V

±12.5 LSB

Each PCI-DAS64/M2/16 is tested at the factory to assure the board’s overall error does not exceed accuracy
limits shown in

.

6-5