Hardware setup, Configuring channels via dip-switches, Caution – Measurement Computing ZonicBook-Medallion rev.5.0 User Manual
Page 11
![background image](https://www.manualsdir.com/files/797841/content/doc011.png)
Hardware Setup
Configuring Channels via DIP-Switches
ZonicBook channel configuration is determined by DIP-switch settings. The default setting is “Single
Ended, A.C. Coupled ICP (Off).” Possible settings are illustrated below.
CAUTION
Each channel must have its associated DIP-switch properly set to achieve the desired
configuration, as determined by your application. Improper settings can damage
your equipment, and will produce erroneous measurements.
To set the switches:
1. Ensure the ZonicBook is disconnected from power, the PC, and all signal lines.
2. Remove ZonicBook’s bottom cover plate.
3. Set each channel’s DIP-switch according to the desired configuration.
Refer to other documentation if needed.
4. Replace the bottom cover plate.
5. For 16-Channel ZonicBook’s only.
(a) After setting the DIP-switches for Channels 1 through 8,
replace the bottom cover plate.
(b) Remove the top cover plate.
(c) Set the DIP-switches for Channels 9 through 16.
(d) Replace the top cover plate.
Default
Channel DIP-Switch Settings
ZonicBook-Medallion
889596
Getting Started, pg. 5