Section 4

IEEE to Parallel Operation

4.4

DIO5

Not Defined - Always "0"

DIO4

Not Defined - Always "0"

DIO3

Not Defined - Always "0"

DIO2

IEEE Input Buffer Full
When this bit is set, it indicates that the Parallel488 may hold off the
controller on subsequent data transfers. The interface may continue
to accept an additional 512 characters.

DIO1

IEEE Input Buffer Empty
When this bit is set, it indicates that the parallel device has accepted
all previous data sent to the Parallel488.

4.4 IEEE 488 Bus Implementation

The Parallel488 implements many of the capabilities defined by the IEEE 488

1978 specification. These are discussed in the following sections. The bus uniline and
multiline commands that the Parallel488 does not support or respond to include:

Remote Enable (REN)
Go to Local (GTL)
Group Execute Trigger (GET)
Local Lockout (LLO)
Take Control (TCT)
Parallel Poll (PP)
Parallel Poll Configure (PPC)
Parallel Poll Unconfigure (PPU)
Parallel Poll Disable (PPD)

4.4.1 My Talk Address (MTA)

The Parallel488 does not support the transfer of data from the parallel port

to the IEEE controller except as it pertains to serial polls. This is due to the
unidirectional data transfer of the parallel interface.