Specifications

17

Safety voltages

Table 7. Safety specifications (Note 3)

Parameter

Conditions

Specification

Channel-to-earth ground isolation

Continuous

250 Vrms, Measurement Category II (Note 4)

Withstand

2,300 Vrms, verified by a 5 sec dielectric withstand test

Channel-to-channel isolation

Continuous

250 Vrms, Measurement Category II (Note 4)

Withstand

1390 Vrms, verified by a 5 sec dielectric withstand test

Note 3:

Connect only voltages that are within the limits specified in this table.

Note 4:

Measurement Category II is for measurements performed on circuits directly connected to the
electrical distribution system. This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example 115 V for US or 230 V for Europe.

Caution! Do not connect the device to signals or use for measurements within Measurement Categories

III or IV.

Screw terminal connectors

Table 8. Screw terminal connector specifications

Connector type

Screw terminal

Screw terminal wiring

16 to 28 AWG copper conductor wire with 7 mm (0.28 in.) of insulation stripped
from the end.

Torque for screw terminals

0.22 to 0.25 N· m (1.95 to 2.21 lb. · in.)

Table 9. Screw terminal assignments

Screw terminal

Signal

0

CH0+ (CH0 IN HI)

1

CH0

– (CH0 IN LO)

0

CH1+ (CH1 IN HI)

1

CH1

– (CH1 IN LO)

0

CH2+ (CH2 IN HI)

1

CH2

– (CH2 IN LO)

0

CH3+ (CH3 IN HI)

1

CH3

– (CH3 IN LO)

Accessory products

Table 10. Screw terminal connector specifications

ACC-102

Two-position detachable screw terminal connector blocks (quantity ten)

ACC-160

Backshell for use with the ACC-102 two-position screw terminal connector blocks. Provides
strain relief and operator protection from high-voltage signals (quantity four)