Ground terminals
The nine ground (
GND
) connections are identical, and provide a common ground for all USB-1024HLS
functions. Refer to the pinout diagram on page 3-2 for the location of the
GND
terminal.
Counter terminal
Pin 20 (
CTR
) is the input to the 32-bit external event counter. Refer to the pinout diagram on page 3-2 for the
location of this terminal. The internal counter increments when the TTL levels transition from low to high. The
counter can count frequencies of up to 1 MHz.
Pull-up/pull-down terminals
You can wire pin 11 (
Port C Pull-up
/
Pull-down
), pin 13 (
Port B Pull-up
/
Pull-down
), and pin 15 (
Port A Pull-
up
/
Pull-down
) to support a pull-up or pull-down connection. These terminals configure the internal 47K
resistors on the USB-1024HLS. Refer to the pinout diagram on page 3-2 for the location of these terminals.
These terminals are configured as open by default.
!
To configure a pull-up connection for a specific port, wire the pull-up/pull-down terminal to a
USB +5 V
terminal.
Figure 3-5. Pull-up connection for Port A
!
To configure a pull-down connection for a specific port, wire the pull-up/pull-down terminal to a
GND
terminal.
Figure 3-6. Pull-down connection for Port A