beautypg.com

Expansion connector – Measurement Computing AI-EXP32 User Manual

Page 10

background image

AI-EXP32 User's Guide

Installing the AI-EXP32

10

Expansion connector

The AI-EXP32 has a 37-pin male DSUB connector that connects directly to the 37-pin female DSUB connector
on a USB-2416 Series device.

Figure 5. DSUB37 expansion connector pin out

Information on signal connections

General information regarding signal connection and configuration is available in the Guide to Signal
Connections (available at

www.mccdaq.com/signals/signals.pdf

).

Associating CJC channels with thermocouple channels

TC channels must immediately follow their associated CJC channels in the channel array. For accurate
thermocouple measurements, associate CJC channels with the TC channels as listed below.

CJC channels

TC channels

CJC6

TC8 through TC11

CJC7

TC12 through TC15

CJC8

TC16 through TC19

CJC9

TC20 through TC23

CJC10

TC24 through TC27

CJC11

TC28 through TC31

When measuring thermocouples, configure the USB-2416 Series device for differential inputs.