3 testing the ethernet connection, Pr-33-s instruction manual – K-Patents PR-33-S User Manual
Page 12

8
PR-33-S instruction manual
Document/Revision No. Rev. 1.4
Effective: May 15, 2015
IP = 192.168.33.1
IP = 192.168.33.2
IP = 192.168.33.3
IP = 192.168.33.100
netmask = 255.255.255.0
PoE
sw
itch
Figure 2.6
A network of PR-33 sensors.
Note: There are no settings for subnet mask, default gateway or name servers in PR-33,
as these settings are not required.
2.3 Testing the Ethernet connection
When you connect the sensor to a powered PoE switch, the indicator lights should light
up. The actual position and function of the lights depend on the switch, but figure 2.7
shows the indicators on a typical PoE switch. There’s one indicator showing that the
instrument is drawing power, another indicator (next to the Ethernet connector) shows
the connection status. In this case a 100 Mbit/s connection is formed between two
instruments at the switch ports 1 and 3 (a sensor per port), and there is a PoE device
connected in these ports, 1 and 3.