Ivoclar Vivadent Bluephase C8 User Manual
Page 3
7
Table of Contents
Safety
Intended use
8
Indication 8
Safety notes
9
Start-up
Handpiece holder
10
Handpiece 11
Operation
Selecting the curing program
and the curing time
12
Cure Memory function
14
Start
Acoustic signals
14
Light intensity
14
Measuring the light intensity
14
Maintenance and cleaning
Housing 15
Light probes
15
Disposal 15
What if …
16
Warranty / Procedure in case of repair
16
Product specifications
17
English
6
5
Dear Customer,
Optimum polymerization is an important
requirement for all light cured materials in
order to consistently produce high quality
restorations. The polymerization light selected
also plays a decisive role in this respect. The-
refore, we would like to thank you for having
purchased Bluephase
®
C8. Bluephase C8 is a
high-quality medical device which has been
designed according to the latest standard of
science and technology in compliance with
the relevant industry standards.
These Instructions for Use will help you safely
to start up the device, make full use of its
capabilities and ensure a long service life.
Should you have any further questions,
please do not hesitate to contact us
(see addresses on the back cover).
Your Ivoclar Vivadent Team
Light intensity
Exposure time for composites
Exposure time for Tetric EvoCeram / IPS Empress Direct
800 mW/cm
2
± 10%
20 s
15 s
1) Applies to a maximum layer thickness of 2 mm and provided that the Instructions for Use of the respective material do not state any other recommendation (might be the case, e.g. with dentin shades)
2) Applies to a maximum layer thickness of 4 mm and provided that the Instructions for Use of the respective material do not state any other recommendation (might be the case, e.g. with dentin shades)
3) Applies to a maximum layer thickness of 3 mm 4) Applies to light-curing (use of the base past only) 5) Applies to dual-curing
HIGH POWER Program
Consistently high light intensity for
the polymerization of restorative and
cementation materials for direct and
indirect restorations.
SOFT START Program
Step-by-step increase of the light
intensity with reduced shrinkage
stress and reduced heat development
for the polymerization of restorative
materials.
LOW POWER Program
Reduced light intensity with reduced
heat development for the polymeriza-
tion of adhesives, liners, and restora-
tive materials in areas near the pulp
when restoring Class V cavities.
650 mW/cm
2
t [s]
0
800 mW/cm
2
t [s]
0
800 mW/cm
2
t [s]
0
5
650 mW/cm
2
Filling materials
Composite
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow / Others
• 4 mm
2)
Tetric EvoCeram Bulk Fill
Tetric Basic White
Compomer
3)
Compoglass F / Compoglass Flow
Indirect restorations / Luting materials
Variolink II Base
4)
/ Variolink Veneer / Variolink N Base
4)
/ Variolink N Clear Veneer / Variolink Ultra
5)
/
Dual Cement
5)
/ Variolink II
5)
Miscellaneous
Helioseal / Helioseal F / Helioseal Clear
Monopaque
MultiCore Flow / Multicore HB
Heliosit Orthodontic
Telio Add-On Flow
Telio Stains
IPS Empress Direct Color
IPS Empress Direct Opaque
Curing time
15 s
20 s
15 s
20 s
40 s
per mm ceramic:
10 s per segment
20 s
30 s
40 s
20 s
20 s
15 s
15 s
30 s
Adhesive
AdheSE / AdheSE One F
ExciTE / ExciTE DSC
Tetric N-Bond / Tetric N-Bond Self-Etch
Heliobond (Syntac)
Temporary materials
Telio CS Inlay/Onlay
Systemp.inlay/onlay
Fermit / Fermit N
Telio CS Link / Systemp.link
Miscellaneous
Heliosit Orthodontic
Vivaglass Liner
Curing time
10 s
10 s
10 s
10 s
20 s per segment
20 s
20 s
Filling materials
Composite
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow /
Tetric EvoCeram / Tetric EvoFlow / Tetric /
Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
Heliomolar / Heliomolar HB / Heliomolar Flow / Others
• 4 mm
2)
Tetric EvoCeram Bulk Fill
Tetric Basic White
Compomer
3)
Compoglass F / Compoglass Flow
Curing time
20 s
25 s
20 s
25 s
40 s
Bluephase
®
C8