Maintenance, cleaning, and diagnosis, 2 cleaning, 3 changing the compression spring – Ivoclar Vivadent SR Ivocap Equipment System User Manual
Page 11: 1 monitoring and maintenance
11
This chapter describes the maintenance and
cleaning procedures which can be carried out
by dental professionals. All other tasks must
be performed by qualified service personnel
at a certified Ivoclar Vivadent Service Center.
7. Maintenance, Cleaning, and Diagnosis
7.2 Cleaning
Metal parts can be cleaned with warm water
and a cleaning brush. After cleaning, the
flask may be treated with Vaseline to
facilitate the removal of plaster residue.
7.3 Changing the Compression
spring
Please take note of the informational sheet
added to the 'spring washers' set.
The Compression spring are to be
changed by a certified Ivoclar Vivadent
Service Center.
Disconnect power and compressed
air connection before maintenance
and cleaning.
7.1 Monitoring and maintenance
The time for these maintenance procedures
depends on the frequency of use and work-
ing habits of the users. For that reason, the
recommended times are only approximates.
What:
Part:
When:
Check all compressed air
connections for correct fit
Compressed air
connections
Before use
Check the components for
mechanical damage
Pressure apparatus
Flask
Clamping frame
Before use
Check the plastic parts of
the flask for correct fit
Clean contaminated
clamping frame (wax), e.g.
by boiling out
clamping frame
in case of
contamination
Plastic parts of the
flask
Before use
The insulation shoulder of the flask
must only be replaced by a certified
Service Center, since remounting
requires a special mounting
assembly and a special material.
Only the correct mounting
procedure ensures sound function
of the flask.