beautypg.com

Maintenance, cleaning, and diagnosis, 2 cleaning, 3 changing the compression spring – Ivoclar Vivadent SR Ivocap Equipment System User Manual

Page 11: 1 monitoring and maintenance

background image

11

This chapter describes the maintenance and
cleaning procedures which can be carried out
by dental professionals. All other tasks must
be performed by qualified service personnel
at a certified Ivoclar Vivadent Service Center.

7. Maintenance, Cleaning, and Diagnosis

7.2 Cleaning

Metal parts can be cleaned with warm water
and a cleaning brush. After cleaning, the
flask may be treated with Vaseline to
facilitate the removal of plaster residue.

7.3 Changing the Compression

spring

Please take note of the informational sheet
added to the 'spring washers' set.

The Compression spring are to be
changed by a certified Ivoclar Vivadent
Service Center.

Disconnect power and compressed

air connection before maintenance
and cleaning.

7.1 Monitoring and maintenance

The time for these maintenance procedures
depends on the frequency of use and work-
ing habits of the users. For that reason, the
recommended times are only approximates.

What:

Part:

When:

Check all compressed air
connections for correct fit

Compressed air
connections

Before use

Check the components for
mechanical damage

Pressure apparatus

Flask

Clamping frame

Before use

Check the plastic parts of
the flask for correct fit

Clean contaminated
clamping frame (wax), e.g.
by boiling out

clamping frame

in case of
contamination

Plastic parts of the
flask

Before use

The insulation shoulder of the flask

must only be replaced by a certified
Service Center, since remounting
requires a special mounting
assembly and a special material.
Only the correct mounting
procedure ensures sound function
of the flask.