What if – Ivoclar Vivadent P710 User Manual
Page 78

78
1902
yes
Data backup – recover
The new software includes program groups which were not yet available at
the time of the data backup. Only the existing program groups of the data
backup were restored.
1911
yes
Data backup – recover
The new software includes program parameters which were not yet
available at the time of the data backup. Please check the individual
program groups. The individual programs are no longer valid.
1912
yes
Data backup – recover
The new software includes program groups which were not yet available at
the time of the data backup. Please check the individual program groups.
The individual program groups are no longer valid.
1920
yes
Data backup – recover
The data backup could not be performed. Please install a newer software
version.
2010
yes
DSA
The tooth region was not determined.
2020
yes
DSA
The 3 shade guide teeth were not found.
2030
yes
DSA
The contour of the tooth was not determined.
2041
yes
DSA
The analysis region and the references differ significantly.
2042
yes
DSA
The analysis region is invalid. Select an analysis region on a tooth surface.
2100
yes
DSA
Unexpected error (# ****)
2754
yes
Error self-test
An error occurred while checking the infrared camera.
2755
yes
Error self-test
An error occurred while checking the infrared camera.
2756
yes
Error self-test
An error occurred while checking the infrared camera.
2757
yes
Error self-test
An error occurred while checking the infrared camera.
2758
yes
Error self-test
An error occurred while checking the infrared camera.
2759
yes
Error self-test
An error occurred while checking the infrared camera.
2760
**
yes
Error IRT
Check the orientation and positioning of the objects. The objects are
possibly too small or are placed in an unfavourable position on the firing
tray. Check if the protective foil has been removed from the sensor or if the
sensor has been contaminated. Check if a correct firing tray is used. The
height of the firing tray must correspond with the height of the firing trays
recommended for the respective materials by Ivoclar Vivadent AG.
2761
**
yes
Error IRT
Check the orientation and positioning of the objects. The objects are
possibly too small or are placed in an unfavourable position on the firing
tray. The object was not sufficiently dried in the specified time. Please
repeat the procedure after the object has cooled down.
2762
**
yes
Error IRT
An error occurred while trying to identify the object. Check if interfering
heat sources are in the visual range of the infrared camera. Sources of
interference are for example: other furnaces, air conditioning systems,
heaters, light bulbs, Bunsen burners, etc. Do not move the objects after the
start of a program.
* Furnace head opens if this error occurs!
** A program in progress is aborted
*** Error cannot be acknowledged; programs cannot be started.
8. What If …